LMP91000: about reading the voltage data of WE #4

Open
chenginhong opened this issue May 20, 2023 · 1 comment
Open

LMP91000: about reading the voltage data of WE #4

chenginhong opened this issue May 20, 2023 · 1 comment

Comments

@chenginhong
Copy link

After I configured it as required in the README (https://github.com/LinnesLab/LMP91000/blob/master/README.md), in the cyclic voltammetry test, I used a multimeter to test the voltage between WE and RE, but I couldn't get the expected data. Can you help me to solve this problem?

@zhangteng1021
Copy link

不接传感器测量,需要将CE短接至RE时,才可以正确测量出RE和WE引脚之间偏差。(来自TI论坛: https://e2echina.ti.com/support/sensors/f/sensors-forum/215680/lmp91000)