Entretien lasersaur #97

Closed
11 of 12 tasks
nicolasdb opened this issue Jun 9, 2018 · 22 comments
Closed
11 of 12 tasks

Entretien lasersaur #97

nicolasdb opened this issue Jun 9, 2018 · 22 comments

Comments

@nicolasdb
Copy link
Owner

nicolasdb commented Jun 9, 2018

ref: https://nicolasdb.github.io/post/lasertech1/

!!! il y avait une liste sur la machine pour les dates de maintenance, disparu.
ha non, c'est à gauche, bien cachée. BAD.
Faut que ce soit visible sinon, on oublie.

Lasersaur

d'abord l'avant, les crasses peuvent être soufflée à l'arrière et on doit tout recommencer.

  • nettoyer le fond pour les débris
  • nettoyer les axes et roulements
  • vérifier la tension des courroies

Ensuite:

  • retirer les plaques arrières
  • vérifier l'état de:
    • nettoyer le tube si besoin
    • resserrer les connecteurs ver extérieurs (air et data et elec)
    • aspirer et nettoyer la boite arrière
  • Calibrer

option

  • check power avec la sonde de puissance

"CHILLER" en bas à droite.

  • vérifier le niveau d'eau. (à l'arrière, y a une jauge, faut que le niveau soit das le vert. Si besoin, rajouter de l'eau distillée!)
  • nettoyer les filtres.

Compresseur air

  • purger le tube
@nicolasdb
Copy link
Owner Author

nicolasdb commented Jun 9, 2018

Check Power

image

avec la sonde à la sortie du tube, 15sec pour tube 100W
note: refroidir la sonde entre chaques mesures, par ex. dans de l'eau et bien l'essuyer avant usage, peut modifier la mesure.

mA Power Readings 09/06/2018
3 12% 0%
4 16% 4%
5 20% 9%
8 32% 28%
10 40% 44%
11 44% 43%
12 48% 43%
13 52% 41%
14 56% 40%
16 64% 40%
18 72% 38%
20 80% 34%
25 100% 28%

image

image

@nicolasdb
Copy link
Owner Author

nicolasdb commented Jun 9, 2018

Bon,
Timetracking, Environ 4h sans se presser et c'est surtout le sondage qui prend du tps. (c'est ma première fois)

Resultat:

  • les mirroirs ne sont donc pas en cause, ni le reste que j'avais changé petit à petit.
  • l'électronique n'est pas en cause non plus, il n'empèche que laserweb serai plus cool que lasaurapp.
  • le power supply non plus, mais elle avait grillée.
  • le test en mise à feu manuelle, avec la sonde en sortie du tube prouve que c'est bien le tube qui est mort.

NEXT; acheter un tube et une alim pour être tranquille.

image

@nicolasdb
Copy link
Owner Author

Aliexpress

https://www.aliexpress.com/item/Reci-S2-90W-CO2-laser-tube-for-CO2-laser-engrave-and-cut-machine-Reci-S2-90w/32655819062.html

  • Cloudray Reci W1 75W CO2 Laser Tube € 300,90 + 238,51 DHL = 539,41€
  • Cloudray Reci W2 90W CO2 Laser Tube € 343,88 + 254,25 DHL = 598,13€
  • Cloudray Reci W4 100W CO2 Laser Tube € 528,72 + 631,32 DHL = 1.160,04€

@nicolasdb
Copy link
Owner Author

@DewiBrunet
Voilà, j'ai donc bien la confirmation qu'il y a une baisse de puissance carabinée. Le tube est presque cuit.
Et on voit bien qu'il y a une nette baisse de puissance avec un pic à 40%.
Ce qui est étrange, c'est que jusqu'à 40%, la puissance est relativement ce qu'on en attend. ça ne déconne qu'au delà de 40%.

J'vais poser la question sur l'internet.

@nicolasdb
Copy link
Owner Author

https://github.com/nortd/driveboardapp/releases/tag/v18.05

On pourrait tenter une upgrade du logiciel lasersaur.
Y a du neuf.

@nicolasdb
Copy link
Owner Author

@nicolasdb
Copy link
Owner Author

Time to pimp the laser

  • step1 unmount panels
  • Check wirering of manual firing solution
  • Unplug everything linked to HT power supply

how the power supply work?

openfab-lab/toolsDocs#36

Old power supply. MYJG100W
J'aimais bien celle là parce qu'il y a un ampèremètre intégré et que le boitier donne plus confiance que l'autre.

image

image

Je change parce que les mecs du shop m'ont conseillé de remplacer le modèle de l'alim qui n'est idéal pour le tube laser choisi. le RECI W2.
Something about the "glow voltage", it will have best result using these together.

Nouveau modèle RECI-DY13
Les connecteurs sont identiques mais pas répartis tout à fait pareil. Le pole 5V est à gauche. Attention au recâblage.
image

How to wire this shit?

image

@nicolasdb
Copy link
Owner Author

nicolasdb commented Jul 7, 2018

transfer to new connecteur with 5V on the left
Using recommended connection 1

1 2 3 4 5 6
5V TH TL WP G IN
symbole type description
5V Output Power output of 5V, as reference for linear pot. (manual fire : power adjust)
TH Input Signal Laser Control Switch , High Level(>=3V) : laser ON ✳️ ,Low Level(<=0.3V) : laser OFF
TL Input Signal Laser Control Switch , High Level(>=3V) : laser OFF ⛔ ,Low Level(<=0.3V) : laser ON ✳️
WP Input Signal Cooling safety Switch , High Level(>=3V) : laser OFF ⛔ ,Low Level(<=0.3V) : laser ON ✳️
G Signal Ground The frame must be connected with laser machine's enclosure and ground properly.
IN Input Signal Max Laser Power control, can use 0-5V Analog Signal Control

@nicolasdb
Copy link
Owner Author

nicolasdb commented Jul 7, 2018

  • Disconnect cable from tube. Red on the left, White on the right.
  • Emptying the cooling system
  • Disconnect water outlet first (- side). Water will continue to get out trough the inlet. (+ side)
    It's best to heat the tube, it will "ramolir" and will be easy to get it off.
  • when no water left. Disconnect the inlet.
  • take out the tube
  • put the new one in place, put ONE layer of the rubber thingy tape each side. (= some confort and same thickness both side)
    Adjust what is loose. And fix the tube with one set of ziptie.
    Just enough to secure it during replacing tubing on inlet and outlet.
    Inlet MUST be bottom, Outlet MUST be on top To naturally help the air bubble to stay in the cooling system.

image

  • Install the power supply.
    image

  • Fill up the cooler with distilled water.

@nicolasdb
Copy link
Owner Author

nicolasdb commented Jul 7, 2018

  • Power on the chiller.
    Check for air bubble in the tube. If there is. Turn off then on again the chiller, it usually get rid of the last bubble.
  • Secure the tube.
  • check again the level of water of the chiller. The level will drop since a part of the water went to the tube. Adjust level until the upper green limit. Just to be full but not to much full.
    image

note: yeah, outlet of the chiller become inlet of the tube, outlet of the tube become inlet of the chiller. That right.

@nicolasdb
Copy link
Owner Author

nicolasdb commented Jul 7, 2018

Install the laser Hour Meter

Wind the black wire's counter around the laser red wire.
https://www.ebay.com/itm/Laser-tube-hour-meter-for-K40-CO2-YAG-from-30W-to-1000W-4-Lifetime-OTAMAT-UK-/282390911473

image

And push the other end into the laser hour meter.
Install the meter were you want.

image

@nicolasdb
Copy link
Owner Author

https://www.ebay.com/itm/RECI-CO2-Laser-Tube-90W-100W-W2-S2-10000hr-Life-DY10-110V-Power-Supply-/182610457816

Operation conditions
• Water cooling: Using purified water; 2 to 5 liter / minute; water temperature: 10℃ - 40℃.
• The operating environment: temperature 2 - 40℃, humidity 10-60%.
• The working current: test current is 29mA. Maximum working current is 29mA. The running current must be kept below 27mA. The life span can reach 8,000 hours if the current is kept below 25mA. The ammeter indicating current must be connected to the negative electrode of the laser tube.
• When it is working under over - high current for long term, the negative pole will appear light yellow, and the life span will be reduced rapidly.
• To protect dust from going into the insulation sheath, please wrap it with plastic film.

@nicolasdb
Copy link
Owner Author

nicolasdb commented Jul 7, 2018

  • connect HT white wire from tube (- pole) to the amp meter on the left pole.
  • connect the right pole (amp meter) to the HT power supply cable.
  • Test to fire the tube, after turning on the power for chiller and the laser PSU.
    Check if the door is closed, it would trigger the safety and nothing will append. Same as the chiller if not powered on.
  • set the PSU amp to 25mA as max power setting.
    image

@nicolasdb
Copy link
Owner Author

nicolasdb commented Jul 7, 2018

YEAH, it burn baby.

As always, Be extra careful with laser PSU,
9000
(actually, it's 40.000V)

It's High Tension power and it could kill you.

don't touch

@nicolasdb
Copy link
Owner Author

Next step:

  • to make some clean svg for these.
    image

  • Then calibrate and test to cut this panel on mdf.

@nicolasdb
Copy link
Owner Author

nicolasdb commented Jul 8, 2018

power et readings 09_06_2018

image

image

@nicolasdb
Copy link
Owner Author

nicolasdb commented Jul 8, 2018

Calibration, ok enough

Test cut des cibles. y en a presque plus.
...\OpenfabNAS\000WORK\3 teams\Openfab.lab\Tools\Lasersaur\New folder\AlignmentTargets.svg

should go to repo toolDocs

@nicolasdb
Copy link
Owner Author

image

La connection est perdue dès qu'on tire le laser. WTF!?
Besoin d'un hard reboot et pas facile pour la reconnection au réseau.

This isn't a matter of software - it's electromagnetic interference.

image

https://groups.google.com/forum/#!searchin/lasersaur/update%7Csort:date/lasersaur/f3VcTaP3a_Q/uTsvtPQ-BwAJ

Ok, donc.
à chaque fois que le laser est mis en marche, il perturbe tellement autour que le système du laser en perd son réseau.

Solution:
J'ai tout mis en dehors de la machine. Et ça marche.

image

@nicolasdb
Copy link
Owner Author

image

@nicolasdb
Copy link
Owner Author

https://nicolasdb.github.io/post/lasertech1/