Some typos in X-ray tubes sources #108

Open
jureyherrera opened this issue May 17, 2022 · 1 comment
Open

Some typos in X-ray tubes sources #108

jureyherrera opened this issue May 17, 2022 · 1 comment

Comments

@jureyherrera
Copy link

There are some typos in the Scripts generated by the X-ray tubes sources:

Kind regards, J

@srio
Copy link
Member

srio commented May 17, 2022

Thanks. I corrected the "Cumylated" typo in the github, but I did not create a new version yet (it is not important).

The label "@?m" should be correct... meaning that "I do not know at what distance". If you read the "Help" you will see a (very old) discussion on that.

It is important to note that the units for flux and spectral density in the x-ray tube applications are not exactly the same than in SR sources, as there are measured for unit of surface at a given distance (known or not). This "label" is not propagated to "power" ... Just to have that in mind!