proc float_to_bytes {val} {
return [format "0x%s" [exec perl -e "print pack('f>', $val)" | xxd -p]]
}
proc bytes_to_float {val_} {
return [exec perl -e "print unpack('f>', pack('H*', '[format "%0x" $val_]'))"]
}
proc hackcheck {} {
global IS_ORIG
if {$IS_ORIG == 1} {
echo "That won't work without uploading hacked firmware!"
return 0
}
return 1
}
proc p {} {
if {[hackcheck] == 0} return
mem2array val 32 0x20001f00 4
set HIGHP_OVERRIDE $val(0)
set LOWP_OVERRIDE $val(1)
set HIGHT_OVERRIDE $val(2)
set LOWT_OVERRIDE $val(3)
#echo [format "override registers: HP=0x%x LP=0x%x HT=0x%x LT=0x%x" $HIGHP_OVERRIDE $LOWP_OVERRIDE $HIGHT_OVERRIDE $LOWT_OVERRIDE]
if {$LOWP_OVERRIDE != 0} {
echo [format "low pressure: %.1f cm-h2O (set by debugger)" [eval bytes_to_float $LOWP_OVERRIDE]]
} else {
echo "low pressure: 4.0 cm-h2O (default)"
}
if {$HIGHP_OVERRIDE != 0} {
echo [format "high pressure: %.1f cm-h2O (set via debugger)" [eval bytes_to_float $HIGHP_OVERRIDE]]
} else {
echo [format "high pressure: %.1f cm-h2O (set by clinician menu)" [eval bytes_to_float [mrw 0x2000e96c]]]
}
set ms_per_breath [expr {.25 * [bytes_to_float [mrw 0x2000e970]]}]
if {$HIGHT_OVERRIDE != 0} {
echo [format "high pressure time: %.1f seconds (set via debugger)" [expr {.000001 * [mrw 0x20001f08]}]]
} else {
echo [format "high pressure time: %.1f seconds (set via clinician menu)" $ms_per_breath]
}
if {$LOWT_OVERRIDE != 0} {
echo [format "low pressure time: %.1f seconds (set via debugger)" [expr {.000001 * [mrw 0x20001f0c]}]]
} else {
echo [format "low pressure time: %.1f seconds (set via clinician menu)" $ms_per_breath]
}
}
proc lp {arg} {
if {[hackcheck] == 0} return
echo "Setting low pressure value to $arg cm-h2O..."
mww 0x20001f04 [float_to_bytes $arg]
p
}
proc hp {arg} {
if {[hackcheck] == 0} return
echo "Setting high pressure value to $arg cm-h2O..."
mww 0x20001f00 [float_to_bytes $arg]
p
}
proc ht {arg} {
if {[hackcheck] == 0} return
echo "Setting high time to $arg seconds..."
set ms_tm [format %d [expr {round($arg * 1000000)}]]
mww 0x20001f08 [format 0x%08x $ms_tm]
p
}
proc lt {arg} {
if {[hackcheck] == 0} return
echo "Setting low time to $arg seconds..."
set ms_tm [format %d [expr {round($arg * 1000000)}]]
mww 0x20001f0c [format 0x%08x $ms_tm]
p
}
proc ra {} {
if {[hackcheck] == 0} return
echo "Clearing setting overrides..."
mwd 0x20001f00 0
mwd 0x20001f08 0
p
}
proc h {} {
echo "Airsense S10 Custom Firmware Debug Commands:"
echo "\tlt \[val\] : set low pressure time interval (seconds)"
echo "\tht \[val\] : set high pressure time interval (seconds)"
echo "\thp \[val\] : set high pressure value (cm-h2O, 0-30))"
echo "\tlp \[val\] : set low pressure value (cm-h2O, 0-30))"
echo "\tra : reset override values to clinician menu settings"
echo "\tp : print all values"
echo "\th : show this help screen"
}