In addition to the DataStorm DAQ development kit running either in standalone mode or remotely via Ethernet cable, it can also run for no-OS software while connected to computer via an usb cable.

Configuration for no-OS software access

The AD7768-1 Evaluation Platform in this configuration is comprised of the following components:

Configure BOOTSEL DIP Switches

Select the SD card as the boot source for the processor.



Configure FMC_VADJ DIP Switches

The FMC_VADJ power rails provide power to the EV-AD7768-1FMCZ via the FMC interface. This is an adjustable voltage. It must be set to 3.3V. Modify the VID_SW DIP switch settings to select 3.3V.

Assemble the Hardware

Follow the steps in the order shown below.

  • Insert the Micro SD card into the SD card slot
  • Insert the EV-AD7768-1FMCZ FMC connector into the FMC mating connector on the Trenz DataStorm DAQ board
  • Connect the Signal Generator to the EV-AD7768-1FMCZ SMA connectors
  • Ensure that Jumper VINSEL is in position B EV-AD7768-1FMCZ board, for it to be powered via the FMC connector
  • Connect the power supply to the DataStorm DAQ
  • Plug the AC-DC adapter into an AC outlet

Signal Source

A signal generator with 2 outputs is used to generate the differential signals driving the AD7768-1 inputs. The AD7768-1 expects the input signals to swing between AVSS and AVDD1 (+5V on the EV-AD7768-1FMCZ board). This means that the user is expected to provide a true differential input centered at VCOM (AVDD1-AVSS)/2 (as in yellow tag #5 below)

Install (if not already done) the Digilent WaveForms App.

Open the WaveForms App and Click on Wavegen in the left margin which opens a waveform window

Click on Channels and select 2
Select Type as Sine
Select Frequency or Period
Set Amplitude to 2V
Set Offset (Common Mode Voltage) to 2.5V
For Channel 2, Match the Channel 1 settings
Select the Phase to 180 degrees for Inverse signal
Click on No synchronization and select Synchronized
Click Run All

Run the no-OS demonstration

Software for no-OS access

There are a few software components that need to be installed in order to run this demo successfully:

  1. A SW terminal like Tera Term VT or Putty

Connect to the target terminal

  • A wired Micro USB serial port connection between the separate Windows PC and the embedded target is required
  • Launch a terminal program like Tera Term VT or Putty and connect via a serial port to interact with the demo
  • Select 115200 baud
  • Select the appropriate target COM port

Run the demonstration

Simply power up the unit. No additional configuration is required. The bare metal application runs automatically and continuously streams analog measurement results across the serial port to be received and displayed in the terminal program running on the separate Windows PC.



Return to AD7768-1 Quick Start Guide